Non-contact Alcohol Concentration Measurement Device At NIR Spectrum

From PC5271 wiki
Jump to navigationJump to search

Team Member

Lim Gin Joe, Sun Weijia, Yan Chengrui, Zhu Junyi

Background

Alcohol (ethanol) is a widely used organic compound in medicine, chemical industry, food, and beverages. Rapid and accurate measurement of alcohol concentration is crucial in various industries, including food and beverage manufacturing, medical disinfection, and industrial synthesis. Existing measurement methods typically have limitations such as complexity, high equipment costs, and large measurement errors. Therefore, developing an efficient, cost-effective, and non-contact measurement method is essential.

Objective

This research proposes a non-contact alcohol concentration measurement method and infrared spectroscopy. This method measures changes in infrared light intensity transmitted through alcohol solutions, using photoelectric sensors and data fitting techniques to achieve high precision measurements.

Schematic Diagram of the Experiment

Advantages of Infrared-Based Alcohol Detection

  • Non-contact measurement, reducing contamination.
  • High accuracy, based on alcohol-specific infrared absorption characteristics.
  • Low cost, with simple photodetection devices.

Equipment Required

  • Oband Light Source: A broadband light source capable of emitting light across a spectrum of wavelengths.
  • Photodetector: A device to measure the intensity of light that has passed through the sample solution.
  • Alcohol Samples: Different alcohol solutions with known concentrations (e.g., ethanol, isopropanol).
  • Cuvette: A transparent container to hold the alcohol solution for light transmission.
  • Spectrometer: For recording and analyzing the intensity of transmitted light.
  • Power Supply: To power the Oband light source.

Methodology

1. Setup the Apparatus

  • Position the Oband Light Source: Place the Oband light source so that the emitted light can be directed towards the alcohol sample in the cuvette.
  • Align the Photodetector: Position the photodetector on the opposite side of the cuvette to measure the intensity of light that passes through the alcohol solution.
  • Cuvette Setup: Fill the cuvette with the alcohol sample. Ensure that the path of light through the cuvette is clear and that the cuvette is free from air bubbles.
  • Ensure Proper Calibration: Before measuring the alcohol samples, calibrate the setup by using a solution of known concentration or distilled water to establish a baseline (zero absorption).

2. Measure Absorbance

Ethanol Sample Preparation
  • Preparation of alcohol solutions at concentrations of 25%, 50%, and 75%.
Ethanol Sample
  • For each sample, measure the light intensity that reaches the photodetector. Record the values using the spectrometer.

3. Analyze the Results

  • Use Origin software for spectral normalization and baseline correction, and process the alcohol spectra at three different concentrations.

Data Analysis

1. Beer-Lambert Law: The relationship between absorbance and concentration is governed by the Beer-Lambert Law, which states:

  A=ϵcl
  Where:
  * A is the absorbance,
  * ϵ is the molar absorptivity (a constant for the alcohol),
  * c is the concentration of the alcohol,
  * l is the path length of light through the sample.