Optical measurement of atmospheric carbon dioxide
Team members
Ta Na: Responsible for proposing and refining ideas, coordinating experimental records, documenting the experimental process, generating graphs, analyzing experimental data, drawing conclusions, making comparisons, preparing appendices, and formatting.
Cao Yuan: Responsible for refining ideas, assisting with experimental operations, writing the experimental principles section, and formatting.
Gao Yihan: Responsible for writing the research background, experimental objectives, experimental advantages and improvements, and referencing.
Qi Kaiyi: Responsible for error analysis.
Chen Yiming: Responsible for writing the research background, experimental objectives, experimental advantages and improvements, and referencing.
Background
Carbon dioxide (CO₂) is a prominent greenhouse gas whose rising concentration plays a critical role in climate change. In the atmosphere, CO₂ molecules absorb Earth’s outgoing infrared radiation in specific wavelength bands, notably around 15,\ 4.3,\ 2.7,\ and\ 2\ \mu m, which causes heat to be trapped and contributes to atmospheric warming[1]. Indeed, empirical climate analyses have identified increasing CO₂ levels as a primary driver of recent global warming[2]. This dual significance of CO₂—as both an essential climate variable and a gas requiring careful monitoring—motivates the development of accurate sensing technologies. Infrared (IR) absorption spectroscopy provides a robust physical principle for detecting gases like CO₂ by exploiting their unique vibrational absorption fingerprints. When IR light at certain wavelengths passes through a gas, molecules that have vibrational modes resonant with those wavelengths will absorb energy, reducing the transmitted light intensity. Polyatomic gases such as CO₂ exhibit strong vibrational-rotational absorption features in the mid-IR region (e.g., near 4.3 μm) that can be measured for quantification [1]. The relationship between absorbed light and gas concentration is governed by the Beer–Lambert law, which states that the intensity of transmitted light decays exponentially with the path length and concentration of the absorbing species[3]. In practice, this means the amount of IR attenuation at a characteristic wavelength is directly related to the CO₂ concentration along the optical path, enabling concentration measurements via absorbance. Non-dispersive infrared (NDIR) sensors build on this principle by using a broadband IR source and a detector with an optical filter to isolate the wavelength of interest, yielding a selective and linear response to the target gas under most conditions[4]. Among the infrared bands of CO₂, the asymmetric-stretch fundamental near 4.3\ µm is particularly suited to sensing. At this wavelength the transition dipole moment is large, so even trace-level concentrations yield a measurable attenuation of the incident beam[4]. Spectroscopic surveys show that the integrated absorption at 4.3\ µm exceeds that of the next-strongest fundamental band at 15\ µm by roughly one order of magnitude, and is about two orders of magnitude stronger than the overtone/combination band around 2.7\ µm[5]. Independent modelling of atmospheric spectra confirms that the 4.3\ µm and 15\ µm bands are, respectively, the first- and second-most intense mid-IR features of\ CO₂[1]. This exceptionally high line strength translates into superior signal-to-noise performance for compact non-dispersive infrared (NDIR) sensors: by isolating a narrow window centered on \approx4.3\ µm, the detector can quantify CO₂ selectively with minimal cross-interference from other gases[4]. The pronounced absorption at this line therefore underpins our decision to base the instrument’s optical filter on the\ 4.3\ µm peak. Purpose This project’s purpose is to develop a portable, low-cost CO₂ concentration measurement device based on infrared absorption spectroscopy. The aim is to design and construct an instrument capable of accurately quantifying CO₂ levels in ambient air by detecting the attenuation of IR radiation at the 4.3 μm absorption band of CO₂. By leveraging the selectivity of this mid-IR spectral feature, the device will provide reliable measurements of CO₂ while remaining compact and energy-efficient, making it suitable for field use or continuous environmental monitoring. To achieve this, the sensor system will employ a broadband thermal infrared source together with an optical band-pass filter and a thermopile detector, following the classic NDIR configuration. The broadband IR emitter (for example, a heated filament) will produce radiation encompassing the\ 4.3 μm wavelength, and a narrow band-pass filter centered at ~4.3 μm will isolate the specific CO₂ absorption band[4]. As the IR beam passes through the sample gas chamber, CO₂ molecules absorb a fraction of the light at this wavelength, and the thermopile detector (sensitive in the mid-IR range) measures the transmitted IR intensity. The use of a thermopile provides a cost-effective and robust means of converting infrared energy to an electrical signal, but it also necessitates strategies to enhance the detection of small absorption-induced changes. Therefore, a lock-in amplifier technique will be integrated into the detection electronics to improve the signal-to-noise ratio. By modulating the IR source (for instance, with a chopper or by driving it with a sinusoidal waveform) and synchronously demodulating the detector output, the system can extract the CO₂ absorption signal from background noise with high fidelity [4]. This approach effectively amplifies the desired signal and filters out low-frequency drift or random noise, as demonstrated in prior CO₂ sensor research[4]. Overall, the purpose of the project is to realize an IR-based\ CO₂ sensing device that combines selectivity, sensitivity, and portability – using inexpensive components and innovative signal processing – to facilitate accurate CO₂ monitoring outside the laboratory setting. Objectives: Accurate quantification of \mathbit{CO}₂ via IR absorption: Develop the device to reliably measure carbon dioxide concentrations, ensuring that the IR absorption at 4.3 μm can be translated into an accurate quantitative reading of CO₂ levels. This entails calibrating the sensor and optimizing its design so that the CO₂ concentration is determined with high precision and accuracy. Verification of Beer–Lambert law relationship[3]: Experimentally validate that the sensor’s response obeys the Beer–Lambert law, i.e. that the measured IR absorbance varies in a predictable manner with CO₂ concentration and optical path length. This objective will confirm the linear (or known non-linear) relationship between concentration and absorption within the operating range, reinforcing the scientific basis of the measurement. Performance evaluation under varied conditions: Rigorously evaluate the sensor’s performance across a range of operating conditions. This includes testing under different environmental conditions and CO₂ levels to assess the device’s stability, repeatability, and accuracy. The goal is to ensure the sensor maintains consistent performance and can distinguish CO₂ concentration changes despite potential external influences or interfering gases. Real-time signal acquisition and environmental compensation: Implement real-time data acquisition and signal processing algorithms that allow the sensor to deliver continuous CO₂ readings with appropriate compensation for environmental factors. This involves correcting for any baseline drifts or systematic errors (for example, due to ambient temperature fluctuations affecting the thermopile or pressure changes affecting density) and ensuring that the output is a stable, real-time indication of CO₂ concentration. By achieving this, the device will be capable of reliable field measurements, providing immediate feedback on CO₂ levels with minimal lag and robust accuracy. Current Limitations and Prospects for Improvement While infrared absorption spectroscopy offers specificity and sensitivity in CO₂ detection, several limitations must be addressed to optimize the performance of portable sensors. Limitations: Cross-Sensitivity: IR-based sensors can exhibit cross-sensitivity to other gases, such as water vapor, which also absorb IR radiation in overlapping spectral regions. This interference can compromise measurement accuracy. Temperature Dependence: Variations in ambient temperature can affect both the IR source and the detector's response, leading to measurement drift. Without proper compensation, this can result in erroneous CO₂ readings. Sensor Drift: Small vibrations or displacements will cause fluctuations in measurement data. Regular maintenance and recalibration are necessary to ensure long-term reliability. Prospects for Improvement: Advanced Optical Filtering: Employing narrowband optical filters with high selectivity can minimize cross-sensitivity by isolating the specific absorption band of CO₂, thereby enhancing measurement accuracy. Temperature Compensation Mechanisms: Integrating temperature sensors and implementing real-time compensation algorithms can mitigate the effects of ambient temperature fluctuations on sensor performance. Self-Calibration Features: Incorporating self-calibration capabilities can address sensor drift and aging, reducing the need for manual recalibration and maintenance.
References: Wei, P. S., Hsieh, Y. C., Chiu, H. H., Yen, D. L., Lee, C., Tsai, Y. C., & Ting, T. C. (2018). Absorption coefficient of carbon dioxide across atmospheric troposphere layer. Heliyon, 4(10), e00785. https://doi.org/10.1016/j.heliyon.2018.e00785 Stips, A., Macias, D., Coughlan, C. et al. On the causal structure between CO₂ and global temperature. Sci Rep 6, 21691 (2016). https://doi.org/10.1038/srep21691 Mamouei, M., Budidha, K., Baishya, N. et al. An empirical investigation of deviations from the Beer–Lambert law in optical estimation of lactate. Sci Rep 11, 13734 (2021). https://doi.org/10.1038/s41598-021-92850-4 T.A.Vincent, B. Urasinska-Wojcik, J.W. Gardner, Development of a Low-cost NDIR System for ppm Detection of Carbon Dioxide in Exhaled Breath Analysis, Procedia Engineering, Volume 120, 2015, Pages 388-391, ISSN 1877-7058, https://doi.org/10.1016/j.proeng.2015.08.648. Wang, Y., Feng, Y., Adamu, A.I. et al. Mid-infrared photoacoustic gas monitoring driven by a gas-filled hollow-core fiber laser. Sci Rep 11, 3512 (2021). https://doi.org/10.1038/s41598-021-83041-2
Background: The escalating concentration of carbon dioxide (CO₂) in Earth's atmosphere is a principal driver of anthropogenic climate change. CO₂, a potent greenhouse gas, absorbs infrared (IR) radiation, particularly at wavelengths around 4.3 μm and 15 \mu m, leading to the trapping of heat and a subsequent rise in global temperatures. Accurate monitoring of CO₂ levels is thus imperative for climate research, environmental monitoring, and the implementation of mitigation strategies. Infrared absorption spectroscopy has emerged as a reliable method for detecting and quantifying CO₂ concentrations. This technique exploits the characteristic absorption bands of CO₂ molecules in the IR region, allowing for non-invasive and real-time measurements. This project not only serves as a practical application of sensor physics principles but also contributes to the broader goal of developing accessible tools for environmental monitoring. The development of portable and accurate CO₂ sensors based on this principle can significantly enhance our ability to monitor air quality in various settings, from industrial environments to indoor spaces, thereby supporting efforts to address climate change and improve air quality management. Purpose: The primary objective of this project is to design and construct a portable, accurate, and cost-effective CO₂ concentration measurement device utilizing infrared absorption spectroscopy. By targeting the strong absorption band of CO₂ at 4.3 μm, the device aims to detect variations in CO₂ levels through the attenuation of IR radiation passing through a gas sample. The system will incorporate a thermal IR source, a gas chamber, and a thermopile detector equipped with an optical bandpass filter centered at the target wavelength. The integration of a low-frequency chopper and a lock-in amplifier will enhance signal detection by mitigating noise and improving sensitivity. This project emphasizes the following specific objectives: Accurate quantification of CO₂ concentration through measurement of the absorption intensity at the selected IR wavelength. Verification of the relationship between optical absorption and gas concentration using a controlled gas chamber and known CO₂ concentrations. Evaluation of system performance under varying experimental conditions. Implementation of real-time signal acquisition and analysis to reduce errors introduced by external disturbances, such as ambient CO₂ fluctuation, temperature drift, and source heating effects. Current Limitations and Prospects for Improvement While infrared absorption spectroscopy offers specificity and sensitivity in CO₂ detection, several limitations must be addressed to optimize the performance of portable sensors. Limitations: Cross-Sensitivity: IR-based sensors can exhibit cross-sensitivity to other gases, such as water vapor, which also absorb IR radiation in overlapping spectral regions. This interference can compromise measurement accuracy. Temperature Dependence: Variations in ambient temperature can affect both the IR source and the detector's response, leading to measurement drift. Without proper compensation, this can result in erroneous CO₂ readings. Sensor Drift: Small vibrations or displacements will cause fluctuations in measurement data. Regular maintenance and recalibration are necessary to ensure long-term reliability. Prospects for Improvement: Advanced Optical Filtering: Employing narrowband optical filters with high selectivity can minimize cross-sensitivity by isolating the specific absorption band of CO₂, thereby enhancing measurement accuracy. Temperature Compensation Mechanisms: Integrating temperature sensors and implementing real-time compensation algorithms can mitigate the effects of ambient temperature fluctuations on sensor performance. Self-Calibration Features: Incorporating self-calibration capabilities can address sensor drift and aging, reducing the need for manual recalibration and maintenance.
Idea
Climate change has already become one of the most important challenges humanity faces. Carbon dioxide, the major greenhouse gas, is 'heating up' the Earth's atmosphere as they have an absorption peak in infrared wavelengths.
This project aims to build a portable and accurate carbon dioxide concentration measurement device.
Possible measurement mechanisms
- Spectroscopy. Carbon dioxide has absorption peaks at 15, 4.3, 2.7 and 2 μm. With a laser/source of such wavelength and the correct filter, we could judge the concentration by absorption levels. See viasala for inspiration of a potential real-time reference design.
- Interferometry. Carbon dioxide has a slightly different refraction index for all wavelengths. By observing the change in interference patterns, we can tell if there is a change in carbon dioxide concentration due to a change in effective path length.
In the end we chose single-wavelength spectroscopy, i.e. absorbance measurement to measure the CO2 content.
Experiment Preparation
Components
- Light source:
- A thermal/blackbody radiation/light source that can emit EM radiation at 4 micrometers. Done with a 12W halogen lamp, mounted on an IKEA SVALLET lamp stand.
- Sensor:
- Thermopile detectors: G-TPCO-035 / TS418-1N426: Thermopile detector with a built-in optical bandpass filter for light around 4μm wavelength for CO2 absorption
- multimeter
- Gas chamber:
- airtight acrylic box surrounding all optical components (chosen)
- metal (stainless steel) extrusion with hollow structures sealed with transparent SiO2 cover glass (chosen)
- gas chamber with specific materials (Si/SiO2) transparent to 4 micrometer wavelength
- Power supply and micellaneous electronics:
- banana wires
- CO2 source:
- bottled CO2
- dry ice
- to look up
- extinction coefficient of CO2
Experiment Principles
The experiment is based on the absorption of CO2 of 4μm-wavelength infrared light. This absorbance is dependent on the extinction coefficient of CO2, which is dependent on its partial pressure, and the distance the light travels. It could be described using the Beer-Lambert Law:
while Fr is the flux density the thermopile detector receives, Fi is the flux density of the source, β is the extinction coefficient and s is the distance.
Yet G-TPCO-035 / TS418-1N426 does not include any kind of flux/temperature/energy to voltage conversion curve in its spec sheet. This means an assumption needs to be made for the flux density-voltage relation of the thermopile detector. We assume the flux density-voltage relation is linear (linear F-V assumaption), i.e. it follows:
while k is a non-zero constant and b is a constant. Both k and b are unknown and not required to be known for our purposes. To overcome this, we designed our general experiment process as follows:
- Pump in CO2 and maintain its stability without turning on the light, and take the first measurement at t=0s.
- Turn on the light, take the measurements for t=3s, 10s, 20s and 30s or else, respectively.
- Switch off the light until properly cooled to room temperature while releasing the CO2 gas back into the room.
- Above entails a single experiment. Repeat measurements.
As a result, two more assumptions are made for this experiment:
- The partial pressure of CO2, thus the extinction coefficient, remains constant during each experiment. (constant β assumption)
- The incident flux density at each moment, i.e. t=0s, 3s etc. remains constant in each day's experiment, assuming the heating of the bulb is regular across experiments. (incident flux assumption)
It follows that within any single experiment, the difference between measurements at different moments should be proportional to the difference between Fr at respective moments, and thus proportional to the exponential term of the Beer-Lambert Law. Mathematically it writes
Then, given the assumptions, , which in turn gives us the expression for the measured extinction coefficient:
As literature tells that βair = 0.535 cm-1, we can figure out the extinction coefficient at any given moment with the presence of CO2 with the help of a reference measurement with air. Hence the partial pressure of CO2 can be described as:
Note that these assumptions are not perfect. Problems such as gas leakage and temperature disturbance can strongly undermine the validity of the constant β assumption, while manual control of the light source, timing and logging of the data may also undermine the accuracy of the incident flux assumption. As experiment setup upgrades, these assumptions prove to be more valid as will be shown in later parts of the page.
Experiment process
....
Experiment 1
In the initial phase of the study, the pre-designed and precision-fabricated acrylic panels were assembled to construct an airtight chamber intended for the containment of carbon dioxide (CO₂) gas. Following assembly, the experimental setup, as illustrated in the accompanying figure, was employed to preliminarily introduce CO₂ into the chamber using methods such as syringe injection and balloon release. However, it was observed that these approaches yielded limited gas inflow, failing to achieve the target concentration levels required for reliable measurements. To overcome this limitation, the methodology was refined by directly placing dry ice (solid CO₂) within the chamber and employing controlled heating to expedite its sublimation, thereby substantially increasing the internal CO₂ concentration. Throughout the process, a high-precision multimeter was utilized to continuously monitor the sensor’s voltage output, which correlates with variations in light intensity caused by changes in CO₂ concentration. For data processing, the Beer-Lambert law, expressed as:
As literature tells that βair = 0.535 cm-1, we can figure out the extinction coefficient at any given moment with the presence of CO2 with the help of a reference measurement with air. Hence the partial pressure of CO2 can be described as:
The recorded voltage values, representative of transmitted light intensity, were subsequently converted to CO₂ partial pressures and corresponding concentrations. The resulting concentration-time profiles were plotted to analyze the gas dynamics within the system.
Experiment setup
....
Data acquisition and analysis
....
Experiment 2
....
Experiment setup
....
Data acquisition and analysis
....
Experiment 3
....
Experiment setup
....
Data acquisition and analysis
....
Comparative studies
....